16 Probes and Biopsy
Operator’s Manual 16 - 23
b. Insert a biopsy needle with the same specification as that of the guiding block into the
hole of the guiding block.
3. Release the needle from the bracket:
a. Remove the guiding block slightly along the direction of the needle’s tail.
b. Separate the residual part of the needle-guide bracket and the probe from the needle.
4. Remove the needle-guided bracket:
Remove the support of needle-guided bracket from the probe.