EasyManua.ls Logo

Mindray Passport 12 - 5.2.2.3 CO2 Calibration

Mindray Passport 12
106 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Passport 8/12 Service Manual 5-3
5.2.2.3 Calibration
Tools required:
A steel gas cylinder with 5±0.03% CO
2
, 21.0% O
2
and balance gas N
2
(P/N 0075-00-0033-01), or a steel gas cylinder
with:
CO
2
concentration 3% - 7%
a/c 0.01 (where a = absolute gas concentration accuracy, c = gas concentration)
balance gas N
2
T-shape connector
Tubing
To perform a calibration:
1. Make sure that the sidestream or microstream CO
2
module is warmed up.
2. Check the airway for leaks.
3. Select [Main Menu]
[Maintenance >>]
[User Maintenance >>]
enter the required password
[Module
Maintenance>>][Maintain CO2 >>]
[Calibrate CO2 >>].
4. In the [Calibrate CO2] menu, select [Zero].
5. After the zero calibration is finished successfully, connect the equipment as follows:
6. Open the valve to flow CO2 and make sure that there is flow sufficient to vent to atmosphere.
7. In the [Calibrate CO2] menu, enter the CO2 concentration in the [CO2] field.
8. In the [Calibrate CO2], the measured CO2 concentration is displayed. After the measured CO2 concentration
becomes stable, select [Calibrate CO2] to calibrate the CO2 module.
If the calibration is finished successfully, the message [Calibration Completed!] is displayed in the [Calibrate CO2]
menu. If the calibration failed, the message [Calibration Failed] is displayed. In this case, perform another calibration.
Monitor
Cylinder
Flowmeter
Relief valve
T-shape connector
Sampling line

Table of Contents

Other manuals for Mindray Passport 12

Related product manuals