Passport 8/12 Service Manual 7-15
2. Unscrew the three M3×8 screws that fasten the microstream CO
2
module, and then remove the microstream
CO
2
module.
7.5.7 Removing AG Module
1. Unscrew the three M3×6 screws and the captive screw that fasten the AG module bracket and connector panel.
Then disconnect all the cables and tubing connecting the connector panel and the AG module to separate the
connector panel and bracket.
2. Unscrew the six M3×6 screws that fasten the AG module and remove AG module.
NOTE
Manage the tubing properly during reassembly and make sure no hoses are kinked or pinched.