Structure and Assembly/Disassembly 9-93
9.3.38.10 Sensor
The disassembly tool: cross-headed screwdriver (M3, M4).
1. See Chapter 9.3.38.6 for details.
2. Unscrew 2 M4 X 12 cross panhead screws on UC-1.5 and UC-0.5 fixing knob with the
screwdriver (M3, M4) to cut off the tie cable CHS-3X100mm from the position signal cable. Pull
the position signal cable from the lift control PCBA assembly.
NOTE
Let the position signal cable through the bales when installing.
3. Unscrew 4 M3 X 8 cross panhead screws from the anti-collision sensor with the screwdriver
(M3, M4) to remove the signal cable.
Fixing knob UC-0.5
Position signal cable
C
able tie CHS-3X100mm
MOT
OR
brushless DC 24V 50 W
reduction ratio
30