Structure and Assembly/Disassembly 9-7
Unscrew 6 M4 X 12 cross panhead screws from the base of the hook management with
cross-headed screwdriver (M3, M4) to remove 3 pothooks.
9.3.4 Mesh of the Base
Pull out the dust-proof covers (For TDK power, there is five dust-proof covers on the left/right side
and front side of the machine; For UR power, six dust-proof covers).
9.3.5 Display (monitor) Assembly
The disassembly tool: cross-headed screwdriver (M3, M4), diagonal cutting pliers.
1. Remove 2 pads from the cable holes.
2. Unscrew 2 M4 X 12 screws with cross-headed screwdriver (M3, M4).