Setup 12-11
Display
Tap [Display] to enter the page, you can set output resolution and range for the connected external
display.
12.1.11 Maintenance
The [Maintenance] function is designed for you to import or export user data, restore factory
setting and export log. You may also execute self-test and option installation/trial through the
maintenance menu. If you require other maintenance functions, please contact Mindray Customer
Service Department or sales representative.
Exporting Setup Data
This function is used to write all setup data of the system into a disk for backup. The format of the
data file is .PDP.
You can select two types of preset data to export from the system:
General module preset data: including “All Preset”, “Image Preset” and “DICOM/HL7”
data.
Exam mode related preset data, including all image setting, comment and body mark
setting and measurement setting data.
Procedures:
1. Select the target module.
2. Tap [Export] to open the [Export Data] screen.
3. Select the path to save the data.
4. Select the exported file and type as PDP and tap [OK].
Importing Setup Data
This function is used to import the existing setup data to the setup data memory of the system. The
system will reset and operate according to the setup preferences that were imported.
Procedures:
1. Tap [Import] to open the Load Data screen.
2. Select the imported file.
3. Tap [OK], a progress bar will appear and the setup data is imported to the specified path.
4. To restore the factory setup data, tap [Load Factory] on the right side of the screen.
You can use [Load Factory] at the bottom of the screen to restore all factory setup data of the
system.
Other Settings
Type
Item
Description
Load Factory
Local factory default settings.
Setup
Show Metrics Logs
Show the metrics logs, such as login count, total usage time,
user metrics details, exam mode metrics details or probes.
Extract Preset Data
/
Export Log
Export the log.
Self-Test
Perform system self-test and restart the machine.
Enter Windows
For password, please contact your service engineer.
Recover
To recover the system.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray TE7T and is the answer not in the manual?

Mindray TE7T Specifications

General IconGeneral
BrandMindray
ModelTE7T
CategoryMedical Equipment
LanguageEnglish

Related product manuals