Probes and Biopsy 13-51
3. Sterilize the needle-guided bracket by using an appropriate method.
Method
Procedures
Sterilant solution
Follow the following steps:
1. Prepare a sterilant by using sterile distilled water when
necessary.
Recommended sterilant: Glutaraldehyde (2.4%), Cidex
Activated Dialdehyde Solution
2. Immerse the needle-guided bracket fully in the sterilant and
shake the bracket appropriately to remove any bubbles on
the surface. Use a syringe to draw an appropriate amount
of sterilant and inject the sterilant into the hole to remove
the bubbles inside the hole if necessary.
For details about the immersion duration, see the
operator’s manual provided by the manufacturer.
3. After sterilization, wash the needle-guided bracket
thoroughly by using a large amount of sterile distilled water
(about 2 gallons) at room temperature for about 30 s to
remove the residual sterilant.
Repeat the operation twice.
4. Dry the needle-guided bracket by using a piece of sterile
disposable lint-free soft cloth.
Sterilizing system
Follow the following steps:
1. Package the needle-guided bracket in accordance with the
sterilization requirements of surgical instruments.
2. Place the packaged needle-guided bracket inside a high-
temperature steam sterilizer and perform sterilization. The
sterilization parameters are 121 °C and 30 minutes for a
gravity displacement steam sterilizer and are 132 °C and 4
minutes for a dynamic-airremoval steam sterilizer.
3. Take out the sterilization package after sterilization and dry
it in an oven at 60 °C for 20 minutes to 30 minutes.
4. Keep the sterilization package together with other sterilized
surgical instruments in a sterile item storage area.
4. Store the needle-guided bracket in a cool, clean and dry environment.
13.2.9 Storage and Transportation
1. Do not use the carrying case for storing the needle-guided bracket. If the carrying case is used
for storage, it may become a source of infection.
2. Between examinations, keep the needle-guided bracket in a sterile environment.
3. If the needle-guided bracket is sent to your MINDRAY representative for repair, be sure to
disinfect or sterilize it and keep it in the carrying case to prevent infection.
4. Sterilize the carrying case as necessary.