System Maintenance 17-3
Dust-proof cover cleaning
1. Press the clip of the dust-proof cover and unfold it outwards (shown in the figure below).
2. To remove the fixing frame of the dust-proof cover and the cover itself, press two clips on the
fixing frame of the dust-proof cover towards arrow’s direction (shown in the figure below).
3. Clean the dust-proof mesh with water or soft brush.
CAUTION:
Please clean all dust-proof covers of the system periodically (1 time
per month); otherwise, system damage may result. Cleaning times
can be increased when the system is used in the open air or
somewhere dust is more.
Cleaning the peripherals
Carry out cleaning maintenance according to your actual peripheral configuration. Items which are
not configured can be skipped.
Content
Description
Color and B/W
video printer
Wipe away dust or stains on the printer cover with a soft dry cloth, then clean
the inside of the printer. Carry out cleaning maintenance according to the
operation manual if necessary.
Graph/text
printer
Wipe away dust or stains on the printer cover with a soft dry cloth, then clean
the inside of the printer. Carry out cleaning maintenance according to the
operation manual if necessary.
Clip
Fixing frame
Clip

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray TE7T and is the answer not in the manual?

Mindray TE7T Specifications

General IconGeneral
BrandMindray
ModelTE7T
CategoryMedical Equipment
LanguageEnglish

Related product manuals