System Maintenance 17-5
Compatible Disinfectants
The following table lists compatible disinfectants.
Manufacturer
Product Name
Type
Advanced Ultrasound Solutions Inc.
SONO
TM
ULTRASOUND WIPES
Wipe
Parker laboratories Inc
Protex
TM
Disinfectant Spray
Spray
PDI Inc.
Sani-Cloth
®
Plus germicidal
disposable cloth
Wipe
Bode Chemie Gmbh
Mikrobac Tissues
Wipe
GAMA Healthcare Ltd.
Clinell universal wipes
Wipe
PDI Inc.
Sani-Cloth
®
AF3
Wipe
Diversey
VIREX II 256
Solution
Diversey
VIREX TB
Solution
Schulke
Schulke mikrozid
®
Sensitive
Wipes
Wipe
/
0.5% sodium hypochlorite
Solution
/
75% alcohol
Solution
Disinfection Procedures
NOTE:
Clean the system in power off status; or in power on status, tap in the top-right
corner of the screen and select to lock the system for 10 seconds.
The ultrasound system has passed puncture test and leakage current test, and thus
can be disinfected on power-on status.
After cleaning, disinfect the main unit as follows.
1. Wear medical gloves to prevent infection.
2. Disinfect the main unit with disinfectant wipe, spray or solution. Follow the disinfectant
manufacturer’s recommended contact time and method.
NOTE: Do not disinfect handle, or any visible sockets or interfaces (such as probe sockets,
ventilation holes, dust-proof cover, loudspeaker, sockets or interfaces in the IO panel and
power supply panel). See the following figure.