Image Optimization 5-17
WF (Wall Filter)
Description
It filters out low-velocity signals to provide effective information, and this function
is used to adjust the filtered frequency. The real-time value is displayed in the
image parameter area in the top-left corner of the screen.
Operations
Adjust through [WF] on the menu.
Select the value according to the actual situation.
Impacts
Flow signals may be missing.
Priority
Description
This function is used to set the levels of the flow display and to display the
grayscale signal or color signal.
Operations
Adjust through [Priority (%)] on the menu.
Tap [-] or [+] to change the level slightly or drag the control directly.
The higher the value, color signals are first to be displayed. The lower the value,
grayscale signals are first to be displayed.
iTouch (Auto Image Optimization)
Description
To optimize image parameters as per the current tissue characteristics for a
better image effect.
Operations
Tap on the left part of the image area to get iTouch optimization.
For probes L12-4s, L9-3s, L11-3VNs, L12-3VNs and L12-3RCs under EM Vas,
Vascular or Carotid exam mode, using iTouch can also achieve ROI optimization
and auto-tracking.
Smart tracking (ROI auto position/steer in Color/PW mode)
Description
To optimize image parameters as per the current tissue characteristics for a
better image effect. When Smart Tracking is turned on, the system optimizes ROI
angle and position automatically to achieve an active tracking by reducing the
impact of patient respiratory movement.
Operation
Turn on/off the function by [Smart Tracking] on the menu.
Impacts
Smart tracking is available only for the following probes under EM Vas, VAS and
Carotid exam mode: L12-4s, L7-3s, L14-6s, L14-6Ns, L16-4Hs, L20-5s 7L4s, L9-
3s, L11-3VNs, L12-3VNs, L12-3RCs and L14-5Ws.
For L12-4s, under B+Color+PW mode, this feature also optimizes PW sampling
line angle, SV size and position.
Dual Live
Description
This function is used to display B image and Color image synchronously.
Operations
Turn on or off the function through the [Dual Live] item on the menu.
When the function is turned on, the window will be automatically switched to the
dual windows (one for B image, and the other for Color image).