5-24 Image Optimization
Adjust through [Trace Smooth] on the menu.
There are 4 levels of smooth effect provided, the bigger the value, the higher the
smooth processing.
This function is used to set the sensitivity of tracing in the spectrum.
Adjust through [Trace Sensitivity] on the menu.
The bigger the value the higher the sensitivity.
In real-time scanning, the results displayed are derived from the calculation of
the latest cardiac cycle.
In the freeze and cine status, the results displayed are calculated from the
current selected area.
Invert
This function is used to set how the spectrum is displayed.
Turn the function on or off using the [Invert] control.
Select “Auto Invert” in the [Setup] -> [System] -> [Image], so the spectrum can
automatically invert when the color flow is steered to a certain angle to
accommodate the operator’s desire to distinguish the flow direction.
Speed
This function is used to set the scanning speed of PW mode imaging.
Adjust through [Speed (mm/s)] on the menu.
Changing the speed makes it easier to identify the cardiac cycles and to detect
more details.
T/F Res.
This function is used to create a balance between time resolution and spatial
resolution.
Adjust through [T/F Res.] on the menu.
Wall Filter
It filters out low-velocity signals to provide effective information, and this function
is used to adjust the filtered frequency. The real-time value is displayed in the
image parameter area in the top-left corner of the screen.
Adjust through [WF] on the menu.
Signals of low-velocity flow may be missing.
Tint Map
This function provides an imaging process based on color difference rather than
gray distinction.
Select maps or turn on/off the function by using [Tint Map] control.