EasyManuals Logo

Mindray uMEC12 Service Manual

Mindray uMEC12
32 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #12 background imageLoading...
Page #12 background image
5. 5.
Select Select
[[
Main MenuMain Menu
] [] [
Maintenance >>Maintenance >>
] [] [
User Maintenance >>User Maintenance >>
] [] [
ModuleModule
Maintenance >>Maintenance >>
] [] [
Cal. IBP Press. >>Cal. IBP Press. >>
]. In the []. In the [
Cal. IBP Press.Cal. IBP Press.
] menu, set the] menu, set the
calibration value to 200 mmHg.calibration value to 200 mmHg.
6. 6.
Select Select
the the
[[
CalibrateCalibrate
] button next to the ] button next to the
desired IBP channel to start a calibration.desired IBP channel to start a calibration.
7. 7.
The The
message message
[[
Calibration Completed!Calibration Completed!
]]
is displayed after a successful calibration.is displayed after a successful calibration.
Otherwise, a corresponding message will be Otherwise, a corresponding message will be
displayed.displayed.
Method 2:Method 2:
Tools required:Tools required:
Standard sStandard s
phygmomanomphygmomanom
etereter
Balloon pumpBalloon pump
TubingTubing
T-shape connectorT-shape connector
1. 1.
Connect Connect
the the
3-way 3-way
stopcock, stopcock,
the the
sphygmomanometesphygmomanomete
r ar a
nd the nd the
balloon balloon
pump pump
through through
aa
T-shape connector, as shown below.T-shape connector, as shown below.
2. 2.
Turn on Turn on
the 3-way the 3-way
stopcock to stopcock to
the air the air
to zero to zero
the transthe trans
ducer, aducer, a
nd then nd then
open the open the
stopcockstopcock
to the to the
sphygmomanometersphygmomanometer
..
3. 3.
Select Select
[[
Main MenuMain Menu
] [] [
Maintenance >>Maintenance >>
] [] [
User Maintenance >>User Maintenance >>
] [] [
ModuleModule
Maintenance >>Maintenance >>
] [] [
Cal. IBP Press. >>Cal. IBP Press. >>
]. In the []. In the [
Cal. IBP Press.Cal. IBP Press.
] menu, set the target] menu, set the target
pressure calibratipressure calibrati
on value of the target con value of the target c
hannel.hannel.
4. 4.
Inflate Inflate
using the using the
balloon balloon
pump pump
until the until the
reading reading
of sphyof sphy
gmomanometer gmomanometer
approximatesapproximates
the preset calibration value.the preset calibration value.
5. 5.
Adjust Adjust
the the
calibration calibration
value value
in in
the the
[[
Maintain IBPMaintain IBP
] menu until it is ] menu until it is
equal to the readingequal to the reading
of sphygmomanometerof sphygmomanometer
6. 6.
Select Select
the the
[[
CalibrateCalibrate
] button next to the ] button next to the
desired IBP channel to start a calibration.desired IBP channel to start a calibration.
SphygmomanometerSphygmomanometer
T-shape connectorT-shape connector
3-way stopcock3-way stopcock
Pressure transducerPressure transducer
Pressure adapter cablePressure adapter cable
IBP moduleIBP module

Other manuals for Mindray uMEC12

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray uMEC12 and is the answer not in the manual?

Mindray uMEC12 Specifications

General IconGeneral
TypePatient Monitor
Display12.1-inch color TFT
ECG3-lead
SpO2Yes
NIBPYes
RespirationYes
TemperatureYes
BatteryRechargeable lithium-ion
NetworkingEthernet
ParametersECG, SpO2, NIBP, Respiration, Temperature
ConnectivityEthernet
Battery LifeUp to 4 hours
Power Supply100-240 VAC, 50/60 Hz

Related product manuals