INVASIVE BLOOD PRESSURES
The V12 or V21 bedside monitors allow a clinician to monitor up to 8 invasive 
blood pressures (IBP) simultaneously. 
Adjust Invasive Blood Pressure Label
To adjust the label on an invasive blood pressure on the V12 or V21 bedside 
monitor:
  a.  Insert IBP Module as needed
  b.   Select the invasive blood pressure waveform to enter IBP Waveform and 
select Setup tab
Or
  b.   Select  Parameters (Setup tab) and select the Invasive Blood Pressure 
(ART, CVP, etc.) to enter the Invasive Blood Pressure’s Setup tab
  c. Select Label 
  d.  Select label
  e.   Select  Accept to save settings and select Done (as needed) to return to 
the main screen 
Note: If the label is assigned to another IBP port, even if the IBP is not being 
displayed or utilized, the label will not be included in the list selection. 
Zero Invasive Blood Pressure
To zero one or more invasive blood pressure transducers on the V12 or V21 
monitor: 
  a.  Insert IBP Module as needed
  b.   Connect the invasive blood pressure cable to the invasive blood pressure 
transducer set
  c.   Position transducer and open the invasive line in accordance with 
hospital policy
  d.  Select  
0
  on the VPS module
Or
  d. Select Zero All* (Quick Functions tab)
Or
*Configurable