h.  Wait 1 minute between injections to allow baseline stabilization
  i.  Perform additional Cardiac Output measurements as needed
  j.   Select measurement to include/exclude in the Cardiac Output average  
(A   in the box indicates cardiac output measurement is selected.  
An   in the box indicates the cardiac output measurement is rejected) 
  k.   Select  Print Runs to print C.O. Run(s) to thermal printer modular recorder
  l.   Select  Hemo Calculations to access the Hemodynamic Calculations 
dialog
Or
  l. Select Done to return to the main screen
Initiate Pulmonary Artery Wedge Pressure (PAWP)
Pulmonary Artery Wedge Pressure (PAWP) is a pressure measurement derived 
from a PA catheter when the PA distal balloon is inflated and the catheter 
advances and occludes a distal pulmonary artery. PAWP pressure is a reflection 
of the pressure in the left ventricle at end-diastole. 
To initiate a PAWP measurement on the V12 or V21 bedside monitor:
  a. Select PA waveform or numeric tile and select PAWP 
Or
  a. Select Procedures tab from the Navigation Area and select PAWP
  b. Select Setup to adjust PA sweep speed, PA scale, or display Art line 
  c. Select Accept to return to PAWP dialog
  d.  Inflate balloon and watch PA waveform for wedge waveform
  e.   Select  Measure to freeze the waveform once a satisfactory wedge 
waveform is detected
  f.  Adjust PAWP reference line to the desired position using Ç or 
È
 
  g.   Select  Accept PAWP to enter the PAWP measurement and Done to return 
to the main screen 
Or 
 g. Resume to repeat measurement
Note: Follow manufacturer’s suggested procedures and hospital policy for 
PAWP balloon inflation.