EasyManuals Logo

Mindray Z50BW User Manual

Mindray Z50BW
237 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #149 background imageLoading...
Page #149 background image
DICOM 10-3
3. Preset local DICOM properties and DICOM server.
Localhost DICOM Service Property
Name
Description
AE Title
Application entity title of the ultrasound system.
The AE title here should be the same with the one of the acceptable
SCU set in the server.
Port
DICOM communication port, which should be the same with the one
in the server.
PDU
Maximum PDU data package size ranging from 16384 to 65536.
The default value is 32768.
Enter AE Title, Port, PDU according to the actual situation.
Note:
AE Title should be the same with the SCU AE Title preset in the server (PACS/ RIS/
HIS), for example, if the AE Title of the server preset in the print server is AAA, and
the AE Title of the accepted SCU is preset as MMM, then in the figure above, the AE
Title of Local should be MMM, and the AE Title of print server should be AAA.
DICOM Server Setting
Name
Description
Device
Name of the device supporting DICOM services.
IP Address
IP address of the server.