EasyManuals Logo

Mindray Z50BW User Manual

Mindray Z50BW
237 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #223 background imageLoading...
Page #223 background image
Electrical Safety Inspection A-3
normal polarity with open neutral(Single Fault Condition),
reverse polarity with open neutral(Single Fault Condition)
LIMITS
For UL60601-1,
300 μA in Normal Condition
1000 μA in Single Fault Condition
For IEC60601-1,
500 μA in Normal Condition
1000 μA in Single Fault Condition
A.6 Enclosure Leakage Test
The following outlet conditions apply when performing the Enclosure Leakage test.
normal polarity( Normal Condition);
reverse polarity( Normal Condition),
normal polarity with open neutral(Single Fault Condition);
reverse polarity with open neutral(Single Fault Condition).
normal polarity with open earth(Single Fault Condition);
reverse polarity with open earth(Single Fault Condition).
LIMITS
For UL60601-1,
100μA in Normal Condition
300 μA in Single Fault Condition
For IEC60601-1:
100μA in Normal Condition
500 μA in Single Fault Condition
A.7 Patient Leakage Current
Patient leakage currents are measured between a selected applied part and mains earth. All
measurements have a true RMS only
The following outlet conditions apply when performing the Patient Leakage Current test.
normal polarity( Normal Condition);
reverse polarity( Normal Condition),
normal polarity with open neutral(Single Fault Condition);
reverse polarity with open neutral(Single Fault Condition).
normal polarity with open earth(Single Fault Condition);
reverse polarity with open earth(Single Fault Condition).

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray Z50BW and is the answer not in the manual?

Mindray Z50BW Specifications

General IconGeneral
BrandMindray
ModelZ50BW
CategoryMedical Equipment
LanguageEnglish

Related product manuals