Safety Precautions 1-9
The probe sheath contains natural rubber that can cause
allergic reactions in some individuals.
Do not use pre-lubricated condoms as a sheath. Lubricant
may not be compatible with the transducer material and
damage may result.
Transducer damage may be caused by inappropriate gel,
detergent or cleanser:
Do not soak or saturate transducers with solutions
containing alcohol, bleach, ammonium chloride
compounds, acetone or formaldehyde.
Avoid contact with solutions or coupling gels containing
mineral oil or lanolin.
Read the following precautions to prevent the probe from malfunction:
Clean and disinfect the transducer before and after each examination.
After the examination, wipe off the ultrasound gel thoroughly.
Otherwise, the ultrasound gel may solidify and the image quality
would be degraded.
To prevent the transducer from being damaged, do not use it where it will be
exposed to:
Direct sunlight or X-rays
Sudden changes in temperature
Dust
Excessive vibration
Heat generators
Use the probes under the following ambient conditions:
Ambient temperature: 0°C ~ 40°C
Relative humidity: 30% to 85% (no condensation)
Atmospheric pressure: 700 hPa ~ 1060 hPa
Use the probe 10L24EA under the following ambient conditions:
Ambient temperature: 0°C ~ 40°C
Relative humidity: 30% to 90% (no condensation)
Atmospheric pressure: 700 hPa ~ 1060 hPa
Use the probe D6-2EA under the following ambient conditions:
Ambient temperature: 18°C ~ 30°C
Relative humidity: 20% to 85% (no condensation)
Atmospheric pressure: 700 hPa ~ 1060 hPa
Repeated disinfection will eventually damage the probe, please check the
probe's performance periodically.