Safety Precautions 1-7
13.
DO NOT turn OFF the power supply of the system during printing, file storage
or invoking other system operations. An interrupted process may not be
completed, and can become lost or corrupted.
14.
The system should be powered by battery when the integrality and reliability
of the protective grounding of external power supply is indeterminate.
15.
The replaceable fuse is inside the chassis. Refer replacing job to Mindray
service engineers or engineers authorized by Mindray only.
Please read the following precautions carefully to ensure the safety of the patient and the
operator when using the probes.
WARNING:
1.
The ultrasonic probe is only for use with the specified
ultrasonic diagnostic system. Please refer to the “2.5.2
Probes Available” to select the proper probe.
2.
The ultrasonic probe must be used only by qualified
professionals.
3.
Confirm that the probe and cable are normal before and
after each examination. Electrical shock may result from
a defective probe.
4.
Do not subject the probe to shock. A defective probe may
cause electric shock to the patient.
5.
Do not disassemble the probe to avoid the possibility of
electric shock.
6.
Never immerse the probe connector into liquids such as
water or disinfectant because
the connector is not
waterproof. Immersion may cause electric shock or
malfunction.
7.
A probe sheath must be installed over the probe before
performing intra-cavity, biopsy or intra-operative
examination.
CAUTION:
1.
When using the probe, wear sterile gloves to prevent
infection.
2.
Be sure to use ultrasound gel. Please use the ultrasound
gel compliant with the relevant local regulations. And
manage the ultrasound gel properly to ensure that it does
not become a source of infection.
3.
In normal diagnostic ultrasound mode, there is no danger
of a normal-temperature burn; however, keeping the probe
on the same region of the patient for a long time may
cause such a burn.
4.
Do not use the carrying case for storing the transducer. If
the carrying case is used for storage, it may become a
source of infection.
5.
It is required to practice ALARA when operating
ultrasound system. Minimize the acoustic powe
r without
compromising the quality of images.

Table of Contents

Other manuals for Mindray Z6

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray Z6 and is the answer not in the manual?

Mindray Z6 Specifications

General IconGeneral
BrandMindray
ModelZ6
CategoryMedical Equipment
LanguageEnglish

Related product manuals