Operation Manual S3000 OPERATING INSTRUCTIONS
7-2
OPERATION AND PROGRAMMING INFORMATION
Power-Up
At power-up, the following message will scroll onto the display:
SONICATOR 3000
MISONIX INCORPORATED
The above screen will stop scrolling, then, after a short time, the MicroTip question will appear. (At any time
during the scrolling message, pressing any key will immediately display the MicroTip question).
Programming the Proper Tip
After the startup screen, the following question will appear on the display:
Are you using a
MicroTip (Y/N)?
The answer to this question sets up proper operating parameters for the horn that will be used. If using a
MicroTip horn, answer “YES” to this question. If using any other type of horn, answer “NO”.
Program Mode or Continuous Operation Selection
Next, the following two screens will alternate on the screen every three seconds:
Press PROG to Program
Timer and PULSAR or
Press START for
Continuous Operation
Pressing START will cause the unit to go into a Continuous mode of operation with the Power setting at 0.0.
(See “Continuous Mode Operation” for the display that will appear). No processing will occur until the Power
Select Knob is turned clockwise to a value greater than 0.0. This mode allows the operator to start
processing quickly without going to any of the programmed memory setups.
To program a setup or access a previously-programmed memory setup, press the PROG key on the keypad.