Operation Manual S3000 OPERATING INSTRUCTIONS
7-3
Programming Mode
When the Programming Mode is selected, the following five screens will cycle in succession, changing at
three-second intervals:
00:00:00 Program 1
00:00.0 00:00.0
Shows currently-programmed timing information for
Program 1 (Program 1 is the default).
No Temperature
Control Selected
This screen may be different depending on
programmed temperature control parameters.
Output Level Set at 0.0
Shows the current program starting output power
knob setting.
Press START to Run,
PROG to Program, ...
Indicates which keypad keys can be used.
Press MENU for Options,
1-9 for Other Programs.
Indicates additional keypad keys that can be used.
For the first screen above, the function of each of the timing values is printed on the display panel next to
the associated value. The top-line timing value represents the programmed Process Time, the left bottom-
line value represents the programmed Pulse-ON time, and right bottom-line value is the programmed
Pulse-OFF time.
For the second screen above, if only monitoring temperature, with no temperature control, the following
screen will appear instead:
Temperature
Monitoring
Also, for the second screen above, if temperature control is programmed, the following screen will appear
instead:
Control Temperatures:
MAX=120
o
C Cool=100
o
C
If no cooling temperature is programmed, “NONE” will
appear in place of that temperature.
At this point, one of several things can be done (detailed information follows):
1) Press any of the nine numeric keys, [1] to [9], to access one of the nine stored Program setups.
2) Press START to begin running using the currently displayed Program information.
3) Press MENU to access other optional setup information.
4) Press CLEAR to return to the MicroTip question.
5) Press PROG to change any or all programmed information values for the current Program that
are displayed.