No. 99MAH054A
5 Setting Parameters
(Short press) / (Long press)
Tips
• If the power is turned ON/OFF using the [F3] key, it is not necessary to reset the
current date. However, if the DC plug is removed, the current date must be reset
when the power is turned on again.
• Enable/disable calibration schedule warning function, current date, calibration date
and advance warning date can be changed.
2
-2
1 2
-1
2
-2
F2F2
F2
F1/F3
Go to step
3
Go to the next item
1 Pressthe[F2]key.
» Calibration schedule warning can be set.
2 Select ON/OFF for the calibration schedule warning.
1 Press the [F1] key or [F3] key.
» Each time the key is pressed, the calibration schedule warning turns ON/OFF
alternately.
2 Press the [F2] key.
If [on] is selected:
» The current date becomes settable and [todAy] blinks.
If [oFF] is selected:
» Selectionisconrmed;shiftstothenextparameteritemof"Otherfunction
changes".
(Go to "5.11.3 Selecting Digimatic Output" on page 67.)