Page 3 of 46
3. Although the EVA Pro complies with EN IEC 60601-1-2:2015 (4
th
ed.) EN
55011 (2009) +A1 (2010) CISPR 11 EN 55011:2016 for electromagnetic
phenomena (EMC) for the Professional healthcare environment, the device
may emit electromagnetic radiation that may affect the performance of
other electrically powered equipment, or the performance of the device
may be affected by electromagnetic radiation from other electrically
powered equipment in the vicinity.
4. Other equipment could interfere with the EVA Pro, even if other equipment
complies with CISPR emissions requirements.
5. NOTE: This equipment has been tested and found to comply with the limits
for a Class A digital device, pursuant to part 15 of the FCC Rules. These limits
are designed to provide reasonable protection against harmful
interference when the equipment is operated in a commercial
environment. This equipment generates, uses, and can radiate radio
frequency energy and, if not installed and used in accordance with the
instruction manual, may cause harmful interference to radio
communications. Operation of this equipment in a residential area is likely
to cause harmful interference in which case the user will be required to
correct the interference at his own expense.
6. The LED light is extremely bright and should be the only illumination used
as a light source during an examination. Do not stare directly into the light
when it is in operation.
7. WARNING: Portable RF communications equipment (including peripherals
such as antenna cables and external antennas) should be used no closer
than 30 cm (12 inches) to any part of the EVA Pro, including cables specified
by the manufacturer. Otherwise, degradation of the performance of this
equipment could result.
8. Only connect the EVA Pro to secured, trusted Wi-Fi networks.
9. The EVA Pro batteries should only be connected to the Provided recharging
base.
10. The EVA Pro device and accessories have not been tested for magnetic
resonance imaging (MRI) safety. Avoid use near MRI equipment.
11. Prior to touching the camera lens or the LED lights, please consult the
manual.
12. The EVA Pro should only be transported in its protective carrying case to
avoid overbalancing.
13. Do not immerse any part of the colposcope in cleaning solutions or any
fluid.