ImageXpress Micro High Content Imaging System Options User Guide
20 5014166 D
Before you run an experiment:
Allow the instrument to warm up for at least two hours.
Make sure the plate sealing ring is at 37°C before use, either by
keeping it inside the instrument or inside an incubator. Incorrect
sealing-ring temperature can cause temperature fluctuation.
Wait 30 minutes to 1 hour for the plate to reach equilibrium. You
can use a journal to set this waiting period.
The current temperature in the chamber is displayed on the
front of the controller.
If there is interstitial space between wells, pipette deionized
water or media into these areas. This helps to increase the
thermal mass of the plate and reduces overall evaporation.
1.
Note: Due to the changes in Z height over time, you might need
to adjust the plate focus parameters in the MetaXpress software
occasionally. To do so, open the MetaXpress software, and then
click Screening > Plate Acquisition Setup > Plate tab > Edit Plate
Bottom Settings. Increase the adjacent well max variation to
accommodate the added Z-height variation introduced by the
fluctuation of temperature.
When setting laser autofocus options, make sure that the
instrument and plate temperatures are at equilibrium. Room-
temperature laser autofocus settings are not ideal when the
system is used at higher temperatures.