8. If applicable, select additional options on the Wavelength tabs.
Wavelength tabs provide options for acquiring individual
wavelength images during a timelapse experiment. See
Figure 1-15 on page 26.
These options include:
At all time points: The default for timelapse experiments. All
time points are acquired for this wavelength.
At start of experiment: Only the first time-point is acquired
for this wavelength.
At start/end of experiment: Only the first and last time
points are acquired for this wavelength.
Every nth time point: Acquires timelapse images only for
every nth time-point for this wavelength.
If you do not image a wavelength at every time point, the most
recent image acquired for that wavelength is saved in the
database for that time-point.