IEC 60601 - testing
level
Electromagnetic environment -
guidelines
Magnetic field in
power supply frequency
(50/60 Hz)
IEC 61000-4-8
Power frequency magnetic fields
should be at levels characteristic of
a typical location in a typical
commercial or hospital environment.
The device contains sensitive
components.
Note: U
T
is the AC mains voltage prior to application of the test level.
RF conducted
disturbance acc. to IEC
61000-4-6
Radiated RF
disturbance
acc. to IEC 61000-4-3
3 Veff
150 kHz to 30 MHz
6 Veff in ISM and
amateur radio
frequency bands
between 150 kHz to
80 MHz
3 Veff
150 kHz to
30 MHz
6 Veff in ISM
and amateur
radio
frequency
bands
between 150
kHz to 80 MHz
Recommended separation distance:
Portable RF
communications
equipment (radio
equipment including its
accessories such as antenna cables
and external antennas) should be
used no closer than 30 cm (or 12
inches) to any parts and cables of
the Vibrasat
®
Pro designated by the
manufacturer. Non-observance may
result in a reduction of the device's
performance.
Radiated RF
disturbance acc. to
IEC 31000-4-3
10 V/m
80 MHz to 2.7 GHz
Table 9 of IEC
60601-1-2 Ed.4
10 V/m
80 MHz to
2.7 GHz
Table 9 of IEC
60601-1-2
Ed.4
Notes:
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection by structures, objects and people.