a) Field strengths from fixed transmitters, such as base stations for radio telephones and land mobile
radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically
with accuracy. To assess the electromagnetic environment due to fixed transmitters, an electromagnetic
site survey should be considered. If the measured field strength at the location where the Vibrasat
®
Pro is
used exceeds the compliance level above, the Vibrasat
®
Pro should be observed to verify normal
operation. If abnormal performance is observed, additional measures may be necessary, such as
reorienting or relocating the Vibrasat
®
Pro.
b) Above the frequency range of 150 kHz to 80 MHz the field strength should be lower than 3 V/m.
The Vibrasat
®
Pro meets all test levels according to IEC60601-1-2 Edition 4
(Tables 4 to 9).
Do not operate the devices of the Vibrasat
®
Pro series directly next
to or stacked with other devices. If operation close to or stacked on
other devices is necessary, observe the devices in the Vibrasat
®
Pro
series to ensure they are operating correctly.
An up-to-date list of the available accessories can be found on our
website www.moeller-medical.com or in our brochure.