4.3 Setup and commissioning
Before commissioning, the Vibrasat
®
Pro Console must be processed as
per the hygiene guidelines (see Chapter 6).
The Vibrasat
®
Pro Wand and the cannula holder must be processed as
per the reprocessing instructions provided by the manufacturer.
If the devices Vibrasat
®
Pro Console and Vibrasat
®
Pro Wand had been
exposed to temperature or humidity fluctuations during transport or other
change of place, they need to rest for at least 2 hours in the operating
environment before restart.
Place the Vibrasat
®
Pro Console on a suitable, stable surface or, if available, use the
Vibrasat
®
Pro fastening kit. Fasten the fastening kit to a standard rail. Put the
Vibrasat
®
Pro Console on the plate and secure it to the fastening kit using the screw
supplied.
Connect the foot switch (optional) to the Vibrasat
®
Pro Console using the connecting
cable.
Insert the mains cable in the designated connector on the Vibrasat
®
Pro Console and in
a socket with a connected earth wire. Observe the voltage indicated on the identification
plate.
Press the On / Off switch on the rear of the Vibrasat
®
Pro Console to switch it to standby
mode.
Remove the sterile Vibrasat
®
Pro Wand from its packaging under sterile working
conditions and connect it to the Vibrasat
®
Pro Console.
Connect the cannula holder to the handle Vibrasat
®
Pro Wand.