Technical Specification/Certification
22
5.2 Certification
China
Ex ia IIC T4 Ga (Ta = -20 °C to +50 °C)
USA (Pending Approval)
Class I, [II,III] Div. 1, Groups A, B, C, D; Ta = -20 °C +50 °C
Canada (Pending Approval)
Class I, Div. 1, Groups A, B, C, D; Ta = -20 °C to +50 °C
Marking, Certificates and Approvals According to the Directive 94/9/EC
[ATEX]
Manufacturer: MSA (China) Safety Equipment Co,Ltd
No.8, Rui En Lane, Xingpu Road Suzhou Industrial Park,
Jiangsu, China 215126
Product: ALTAIR Pump Probe
EC-Type Examination
Certificate:
Type of protection: EN 60079-0:2012, EN 60079-11:2012
Performance: none
Marking:
II 2G Ex ia IIC T4 Gb
-20 °C ≤ Ta ≤ +50 °C
Li-Ion: Um = 6,7 V
Special Conditions: The equipment shall be charged only in no-hazardous
locations by manufacturer`s chargers only. The charging
voltage shall not exceed 6.7 V.
Quality Assurance
Notification:
0158
Year of Manufacture: see Label
Serial Nr.: see Label
FASIT Srl - Via Orsera 13/b - 34145 TRIESTE - ITALY - Tel +39 040 825604 - www.fasit.it - info@fasit.it