Operation
16
3.6 Battery Charging
The ALTAIR Pump Probe uses a non-user serviceable rechargeable Lithium
Ion battery.
The charger is capable of charging a completely depleted pack in less than four
hours in normal, room-temperature environments.
- Minimum and maximum ambient temperature to charge the device is 10 °C
[50° F] and 35 °C [95° F], respectively.
- For best results, charge the device at room temperature 23 °C [73,4° F].
To Charge the Device
(1) Firmly insert the charger connector into the charge port on the back of the
device.
(2) The Charge LED is used to indicate the charge status.
Red = Charging
Green = Charged
Orange = Fault
(3) If a problem is detected during charging [Charge LED turns Orange]:
Disconnect the charger momentarilty to reset the charge cycle.
WARNING
Risk of explosion: Do not recharge device in hazardous area.
WARNING
Use of any charger, other than a MSA qualified charging device, may damage
or improperly charge the batteries.
For users in Australia/New Zealand: The charge cradle is a Class A
product. In a domestic environment, this product may cause radio
interferece, in which case, the user may be required to take
adequate measures.
Allow extremely hot or cold devices outside the charging tempera-
ture range to stabilize for one hour at room temperature before
attempting to charge.