Command 133: Set Alarm Relay Action
Set the action/mode of the Alarm Relay.
Request Data Bytes:
Byte Format Description
0 Unsigned-8
0=Latching (default)
1=Sentry
2=Non-Latching
Command Specific Response Data Bytes:
Byte Format Description
0 Unsigned-8 Response Code
1 Unsigned-8 Device Status
2 Unsigned-8 Returns new value of Alarm Relay Action Variable
Command 134: Set Warning Relay Action
Set the action/mode of the Warning Relay.
Request Data Bytes:
Byte Format Description
0 Unsigned-8
0=Latching
1=Sentry
2=Non-Latching (default)
Response Data Bytes:
Byte Format Description
0 Unsigned-8 Response Code
1 Unsigned-8 Device Status
2 Unsigned-8 Returns new value of Warning Relay Action Variable
56 SMC 5100-XX-IT Toxic Gas Detector Module