Command 135: Reset Alarms
This command will reset any alarm relays that have been latched, providing the alarm condition is no longer present on the
sensor.
Request Data Bytes: None
Response Data Bytes:
Byte Format Description
0 Unsigned-8 Response Code
1 Unsigned-8 Device Status
Command 136: Abort Calibration
This command aborts the calibration procedure.
Request Data Bytes:None
Response Data Bytes:
Byte Format Description
0 Unsigned-8 Response Code
1 Unsigned-8 Device Status
Command 137: Set Calibration Gas Level
Request Data Bytes:
Byte Format Description
0 Unsigned-8 Calibration Gas Level, range 0 to 100
Command Specific Response Data Bytes:
Byte Format Description
0 Unsigned-8 Returns new Calibration Gas Level
Command 138: Apply ZERO Gas
This command will signal the module to accept the present sensor output to as the ZERO gas condition.
Request Data Bytes: None
Response Data Bytes:
Byte Format Description
0 Unsigned-8 Response Code
1 Unsigned-8 Device Status
SMC 5100-XX-IT Toxic Gas Detector Module 57
US 13 HART

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the MSA SMC 5100-IT Series and is the answer not in the manual?

MSA SMC 5100-IT Series Specifications

General IconGeneral
BrandMSA
ModelSMC 5100-IT Series
CategoryGas Detectors
LanguageEnglish

Related product manuals