34
Figure 11-1
Test Point (TP) Locations
Figure 11-2
AOM Signals
A
B
C
D
E
F
G
H
I
Test Point #9: 0 to 10 Volts as the magnet moves back and forth.
Test Point #12: 0 to10 Volts or ±10 volts as the magnet moves back and forth.
(The 4-20 mA Ungrounded option is 0 to 2 Volts)
TB1-A: Same as H, above.
5 Volts at Test Point 1
0 Volts, TB3-K
3 - 5 Volt Interrogation
pulse sent to TB2-E
Pulse returned from TB2-C
(approximately 2-3 Volts)
TTL level pulse, 3-5 Volts
located at Test Point 3
Pulse-width modulated output
3-5 Volts at Test Point 2
6 Volts at Test Point 6
1 ms
20ms 15ms 20ms
Update
Time
3
9
7
6
11
8
12
1
2
10
5
4