IMC01 Maintenance Manual
Version 1.03
Figure 4-14 Start the CIS Calibration
6) After finish, just take the calibration paper out and close the covers, and then turn off the machine.
7) Turn on the machine to finish the CIS calibration.
4.3. Back to Default Setting
After fine tuning the screw and CIS calibration, if there are still some errors or the counting is still not correct,
returning to the factory default setting mode is required.
1) Enter the menu interface, and go to the service menu with the password “9999”, as shown in the following
figures.
Figure 4-15 Service Menu
2) As shown in above figure, choose “. Back to Default Setting” and enter the menu key. It will enter to the
following display on the screen.
Figure 4-16 Default Setting Interface
3) Press the RESTART key, the IMC01 will reset all the changed settings before.
Figure 4-17 Default Setting Guide