TECHNICAL MANUAL - hyperion X5
EN
CONTROL BOARDS DESCRIPTION
45
LED ID COLOR NAME ON BOARD DESCRIPTION
DL9 YELLOW VDC
On when 390V are present on power amplifier.
Off when in stand-by or in case of issues with the power amplifier.
DL10 GREEN -5VA Analog -5.3V present, normally on
DL11 GREEN +5VA Analog +5.3V present, normally on
DL13 GREEN 18V 18V driver tension present, normally on
DL14 GREEN 12V 12V driver tension present, normally on
DL15 GREEN 5.0V 5V CPU tension present, normally on
DL16 GREEN 12V_EXT 12V for x-ray button, emergency button and interlock, normally on.
DL17 GREEN 3.3V 3.3V CPU tension present, normally on
DL18 GREEN 24V 24V tension present, normally on
DL19 GREEN SYNC Power-LOGIC boards sync signal, normally off
DL20 GREEN ENB
Inverter enabled signal from LOGIC board, normally on. Off during standby or
in error condition.
DL21 GREEN ACK Acknowledged signal from LOGIC board, normally off.
DL23 GREEN RDYL External “ready” lamp relay status. Normally off, on when unit is “ready”.
DL24 YELLOW XRYL External “x-ray on” lamp relay status. Normally off, on during x-ray emission.
DL25 GREEN BUT Xray button pressed. On when x-ray button is pressed.
DL26 YELLOW ILK Interlock status. Normally off. On if door is open or interlock is not connected.
DL27 RED EMG
External emergency button status. Normally off. On if External emergency
button status is pressed or not connected.
DL28 YELLOW FIL
X-ray filament on. On during emission.
DL29 YELLOW INV Inverter on. On during emission
DL30 YELLOW DRV Driver enabled, normally on. Off during standby or in error condition.
DL31 GREEN 48V 48V tension ON. Off during standby or in error condition.
DL32 YELLOW ON
Mains power supply to power ampifier relay status. Off during standby or in
error condition.
DL33 YELLOW PC Preload resistance by-pass relay. Off during standby or in error condition.
DL35 RED EMG
Integrated emergency button status. Normally off. On if integrated
emergency button status is pressed or not connected.
DL36 RED MT
Column engine overload. Normally off. On when thermal sensor is active or
not connected.
To connect the X-ray button, use k10 connector as follows:
PIN 1-2: X-ray emission contact
PIN 3: signal "Ready"
PIN 4: signal "X-ray ON"