TECHNICAL MANUAL - hyperion X5
34
CALIBRATION EN
2.4. VERIFYING THE SENSOR CALIBRATION
Click on “Start procedure”, make sure to have
positioned the aluminum filter on the front cover of
the generator and to have removed every object
from the scan area (bite support, chin support,
calibration phantom)
The unit will automatically emit x-rays, performing
26 exposures. At the end of the procedure, perss
“OK” to complete this step.