ELT-400 7 Remote control
7-11
SET:MAX_HOLD<SP><param>[CR]<LF>
SET:MAX_HOLD?[CR]<LF>
Response:
<param><CR><LF>
Meaning:
Activates / deactivates the MAX HOLD display function.
param Explanation
ON Activate MAX HOLD display function
The highest value measured since the function was activated
is output instead of the current measurement value.
The status of the overload display takes into account the
instrument status since the function was activated.
ª Remote control command CALC:OVLD?, see page 7-18
OFF Deactivate MAX HOLD display function
The current measurement value is output instead of the
highest measured value.
The function is deactivated automatically for particular
actions.
ª Default values: see page A-1
Table 7-8 Display function settings