2.4 Dangers from electromagnetic fields
Narda NBM-520 9
2.4 Dangers from electromagnetic fields
Strong Fields
Very high field strengths can occur in the vicinity of
some radiation sources.
Do not cross or ignore safety barriers or markings.
Persons with electronic implants (e.g. heart pacemak-
ers) must avoid danger zones.
Measurement Errors
Metallic labels (stickers) affixed to the yellow sensor
area of the probe can lead to measurement errors, usu-
ally an underestimation of the electromagnetic field
strength.
Affix labels of any type only to the black probe shaft.
If the instrument malfunctions, take it out of service and
contact your Narda Service Center. The addresses are
listed at the end of this operating manual and on the In-
ternet at http://www.narda-sts.com.
Probe is not operating properly
Possibly present high radiation values are not recog-
nized.
Check probes for proper operation with a signal source
before using this measuring instrument. This is espe-
cially important for thermocouples because the sensors
can be affected by various mechanical and environmen-
tal stressors.
Narda offers portable sources to accomplish this import-
ant step (see “Accessories” on page 57).
Before beginning any RF radiation measurement, al-
ways inform yourself of the frequencies and field
strengths that you could expect to encounter.