EasyManua.ls Logo

NARDA nbm-520 - PC Software

NARDA nbm-520
68 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4.8 Selecting the units
Narda NBM-520 33
3. Press the Hold button again for each further
measurement you want to make.
The average value of all the measurements made so
far is displayed,
#... indicates the number of measurements made so
far.
4. If necessary press the Clear button to reset the average
value.
To measure values continuously
You have set the Measurement method to Continuous
using the PC software.
1. Press the Mode button repeatedly until the result type
SPATIAL is displayed.
2. Press the Hold button to start the measurement.
The elapsed measurement time is shown top left in
the display.
3. Move the NBM-520 smoothly through the volume to be
measured and then press the Hold button again to end
the measurement.
The spatial average value is shown on the display.
4. To perform a further measurement, press the Hold
button again. This automatically resets the last average
value.
Note: To help you move the NBM-520 smoothly, an audible signal
is output once every second.
4.8 Selecting the units
The NBM-520 can display the results in several different
measurement units.
The following units can be displayed for normal (flat)
probes:
•V/m
•A/m
•W/m
2
•mW/cm
2
Downloaded from Arrow.com.