EasyManua.ls Logo

National Instruments NI USB-621x - Counter N Gate Signal; Routing a Signal to Counter N Gate; Routing Counter N Gate to an Output Terminal; Counter N Aux Signal

National Instruments NI USB-621x
178 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 9 Counters
© National Instruments Corporation 9-27 NI USB-621x User Manual
Counter n Gate Signal
The Counter n Gate signal can perform many different operations
depending on the application including starting and stopping the counter,
and saving the counter contents.
Routing a Signal to Counter n Gate
Each counter has independent input selectors for the Counter n Gate signal.
Any of the following signals can be routed to the Counter n Gate input.
PFI <0..3>, PFI <8..11>
ai/ReferenceTrigger
ai/StartTrigger
ai/SampleClock
ai/ConvertClock
ao/SampleClock
In addition, Counter 1 Internal Output or Counter 1 Source can be routed to
Counter 0 Gate. Counter 0 Internal Output or Counter 0 Source can be
routed to Counter 1 Gate.
Some of these options may not be available in some driver software.
Routing Counter n Gate to an Output Terminal
You can route Counter n Gate out to any PFI <4..7> or PFI <12..15>
terminal.
Counter n Aux Signal
The Counter n Aux signal indicates the first edge in a two-signal
edge-separation measurement.
Routing a Signal to Counter n Aux
Each counter has independent input selectors for the Counter n Aux signal.
Any of the following signals can be routed to the Counter n Aux input.
PFI <0..3>, PFI <8..11>
ai/ReferenceTrigger
ai/StartTrigger
In addition, Counter 1 Internal Output, Counter 1 Gate, Counter 1 Source,
or Counter 0 Gate can be routed to Counter 0 Aux. Counter 0 Internal

Table of Contents

Related product manuals