Chapter 5 Connecting AI Signals on the USB-6210/6211 Devices
NI USB-621x User Manual 5-10 ni.com
The difference in ground potential between two instruments connected to
the same building power system is typically between 1 and 100 mV, but the
difference can be much higher if power distribution circuits are improperly
connected. If a grounded signal source is incorrectly measured, this
difference can appear as measurement error. Follow the connection
instructions for grounded signal sources to eliminate this ground potential
difference from the measured signal.
When to Use Differential Connections with Ground-Referenced Signal
Sources
Use DIFF input connections for any channel that meets any of the following
conditions:
The input signal is low level (less than 1 V).
The leads connecting the signal to the device are greater than
3m(10ft).
The input signal requires a separate ground-reference point or return
signal.
The signal leads travel through noisy environments.
Two analog input channels, AI+ and AI–, are available.
DIFF signal connections reduce noise pickup and increase common-mode
noise rejection. DIFF signal connections also allow input signals to float
within the common-mode limits of the NI-PGIA.
Refer to the Using Differential Connections for Ground-Referenced Signal
Sources section for more information about differential connections.
When to Use Non-Referenced Single-Ended (NRSE) Connections with
Ground-Referenced Signal Sources
Only use non-referenced single-ended input connections if the input signal
meets the following conditions.
The input signal is high-level (greater than 1 V).
The leads connecting the signal to the device are less than 3 m (10 ft).
The input signal can share a common reference point with other
signals.
DIFF input connections are recommended for greater signal integrity for
any input signal that does not meet the preceding conditions.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the National Instruments NI USB-621x and is the answer not in the manual?

National Instruments NI USB-621x Specifications

General IconGeneral
BrandNational Instruments
ModelNI USB-621x
CategoryComputer Hardware
LanguageEnglish

Related product manuals