© National Instruments | 4-7
X Series User Manual
MIO X Series devices are designed to have fast settling times. However, several factors can 
increase the settling time which decreases the accuracy of your measurements. To ensure fast 
settling times, you should do the following (in order of importance):
1. Use Low Impedance Sources—To ensure fast settling times, your signal sources should 
have an impedance of <1 kΩ. Large source impedances increase the settling time of the 
NI-PGIA, and so decrease the accuracy at fast scanning rates.
Settling times increase when scanning high-impedance signals due to a phenomenon called 
charge injection. Multiplexers contain switches, usually made of switched capacitors. 
When one of the channels, for example channel 0, is selected in a multiplexer, those 
capacitors accumulate charge. When the next channel, for example channel 1, is selected, 
the accumulated charge leaks backward through channel 1. If the output impedance of the 
source connected to channel 1 is high enough, the resulting reading of channel 1 can be 
partially affected by the voltage on channel 0. This effect is referred to as ghosting.
If your source impedance is high, you can decrease the scan rate to allow the NI-PGIA more 
time to settle. Another option is to use a voltage follower circuit external to your DAQ 
device to decrease the impedance seen by the DAQ device. Refer to the document, 
Eliminate Ghosting on Adjacent Input Channels by Decreasing Source Impedance, 
by going to ni.com/info and entering the Info Code rdbbis.
2. Use Short High-Quality Cabling—Using short high-quality cables can minimize several 
effects that degrade accuracy including crosstalk, transmission line effects, and noise. The 
capacitance of the cable can also increase the settling time.
National Instruments recommends using individually shielded, twisted-pair wires that are 
2 m or less to connect AI signals to the device. Refer to the 
Connecting Analog Input 
Signals
 section for more information.
3. Carefully Choose the Channel Scanning Order 
• Avoid Switching from a Large to a Small Input Range—Switching from a channel 
with a large input range to a channel with a small input range can greatly increase the 
settling time.
Suppose a 4 V signal is connected to channel 0 and a 1 mV signal is connected to 
channel 1. The input range for channel 0 is -10 V to 10 V and the input range of 
channel 1 is -200 mV to 200 mV.
When the multiplexer switches from channel 0 to channel 1, the input to the NI-PGIA 
switches from 4 V to 1 mV. The approximately 4 V step from 4 V to 1 mV is 1,000% 
of the new full-scale range. For a 16-bit device to settle within 0.0015% (15 ppm or 
1 LSB) of the ±200 mV full-scale range on channel 1, the input circuitry must settle 
to within 0.000031% (0.31 ppm or 1/50 LSB) of the ±10 V range. Some devices can 
take many microseconds for the circuitry to settle this much.
To avoid this effect, you should arrange your channel scanning order so that transitions 
from large to small input ranges are infrequent.
In general, you do not need this extra settling time when the NI-PGIA is switching 
from a small input range to a larger input range.