Appendix B – Specifications
External weak-acid galvanic oxygen fuel-cell
0.75" 7-segment LED digital display, 4 characters
Displays oxygen from 0 to 100 percent.
0.00–0.99 %
1.00–9.99 %
10.0–99.9 %
100.0 %
Color -Coded LED’s for system status:
Non-isolated 4–20 mA, 12 VDC, negative
ground, powered by analyzer, maximum
electrical load 250 Ohms
0 %–1 %
0 %–10 %
0 %–25 %
0 %–50 %
0 %–100 %
5.63 VDC + 0.1VDC
6.25 VDC + 0.1VDC
6.88 VDC + 0.1VDC
7.50 VDC + 0.1VDC
8.13 VDC + 0.1VDC
Field Adjustable Form C (SPDT) Voltage-free, 5A
@ 250 VAC, 5A @ 30 VDC. Configurable to fail-
safe/non fail-safe and ascending/descending action
Field Adjustable Form C (SPDT) Voltage-free, 5A
@ 250 VAC, 5A @ 30 VDC. Configurable to fail-
safe/non fail-safe and ascending/descending action
Non-adjustable Form B (SPST) Voltage-free, 5A @
250 VAC, 5A @ 30 VDC, non fail-safe action, non-
configurable.
Specifications are subject to change without notice.