Model 1100 Operations Manual
Manual Part Number:
C5-06-4900-01-0
Manual File Name:
MN-A-0004
Manual Revision Level: E
ECO: 9760
Page 46
Appendix B Specifications
OXYGEN SENSOR
External weak-acid galvanic oxygen fuel-cell
DISPLAY
0.75" 7-segment LED digital display, 4 characters
Displays oxygen from 0 to 100 percent.
Resolution:
0.000.99 %
1.009.99 %
10.099.9 %
100.0 %
X.XX
X.XX
XX.X
XXX.X
Color -Coded LED’s for system status:
RUN:
Green
FAULT:
Yellow
ALARM-1:
Red
ALARM-2:
Red
SIGNAL
INTERFACE
Serial Service Port:
Bi-directional RS-232
Analog Voltage Output:
0–1, 05, or 010 VDC
Analog Current Output:
Non-isolated 420 mA, 12 VDC, negative
ground, powered by analyzer, maximum
electrical load 250 Ohms
Range ID Voltage:
0 %1 %
0 %10 %
0 %25 %
0 %50 %
0 %100 %
5.63 VDC + 0.1VDC
6.25 VDC + 0.1VDC
6.88 VDC + 0.1VDC
7.50 VDC + 0.1VDC
8.13 VDC + 0.1VDC
RELAY OUTPUTS
Alarm-1:
Field Adjustable Form C (SPDT) Voltage-free, 5A
@ 250 VAC, 5A @ 30 VDC. Configurable to fail-
safe/non fail-safe and ascending/descending action
Alarm-2:
Field Adjustable Form C (SPDT) Voltage-free, 5A
@ 250 VAC, 5A @ 30 VDC. Configurable to fail-
safe/non fail-safe and ascending/descending action
Fault:
Non-adjustable Form B (SPST) Voltage-free, 5A @
250 VAC, 5A @ 30 VDC, non fail-safe action, non-
configurable.
Specifications are subject to change without notice.