EasyManua.ls Logo

Neutronics 1100 - 5.4 Range and Output Mapping; 5.5 Zero Calibration Settings

Neutronics 1100
60 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Model 1100 Operations Manual
Manual Part Number:
C5-06-4900-01-0
Manual File Name:
MN-A-0004
Manual Revision Level: E
ECO: 9760
Page 50
5.4 Appendix E Range / Output
RANGE NAME
MEASURED
RANGE
DISPLAY
ANALOG
RANGE
RANGE ID
VOLTAGE
OUTPUT
FULL AUTO
RANGE
0.00 % .9 %
1.00 % 9.99 %
10.0 % 24.9 %
25.0 % 49.0 %
50.0 % 99.9 %
100.0 %
X.XX
X.XX
XX.X
XX.X
XX.X
XXX.X
0–1 %
0–10 %
0–25 %
0–50 %
0–100 %
5.63 VDC
6.25 VDC
6.88 VDC
7.50 VDC
8.13 VDC
LOW AUTO
RANGE
0.00 % .99 %
1.00 % 9.99 %
10.0 % 24.9 %
X.XX
X.XX
XX.X
0–1 %
0–10 %
0–25 %
5.63 VDC
6.25 VDC
6.88 VDC
HIGH AUTO
RANGE
10.0 % 24.9 %
25.0 % 49.0 %
50.0 % 100.0 %
XX.X
XX.X
XXX.X
0–25 %
0–50 %
0–100 %
6.88 VDC
7.50 VDC
8.13 VDC
FIXED RANGE
0–1 %
0.00 % .99 %
X.XX
0–1 %
5.63 VDC
FIXED RANGE
0–10 %
0.00 % 9.99 %
X.XX
0–10 %
6.25 VDC
FIXED RANGE
0–25 %
0.0 % 9.99 %
10.0 % 24.9 %
X.XX
XX.X
0–25 %
6.88 VDC
FIXED RANGE
0–50 %
0.0 % 9.99 %
10.0 % 49.9 %
X.XX
XX.X
0–50 %
7.50 VDC
FIXED RANGE
0–100 %
0.0 % 99.9 %
100.0 %
XX.X
XXX.X
0–100 %
8.13 VDC
5.5 Appendix F – Zero Calibration Range Settings
ENTERED
VALUE
RANGE
AVAILABLE
SETTINGS
0
Single Point Calibration
1
1 PPB 50 PPB
2
1 PPB 500 PPB
3
1 PPM 5 PPM
4
10 PPM 50 PPM
5
10 PPM 500 PPM
6
0.0 % 0.5 %
7
1 % 5 %
X
8
1 % 50 %
9
18 % - 24 %
10
10 PPM 20 PPM

Table of Contents