ChipSHOUTER Users Manual: Serial Interface
58
arm a
Arms device (charges high voltage capacitor bank).
If no trigger occurs the device will automatically
disarm after arm_timeout seconds.
If arming fails, the device may have an active
fault. Check active and latched faults with the
get fault command.
Arm A
Clears latched faults and arms device, equivalent
to running set fault none followed by arm. This
command is useful when using the external trigger,
as you may need to quickly clear a latched fault
and arm the device.
disarm d
Disarms device (turns off high voltage and dis-
charges capacitor bank internally).
safe S
See disarm.
pulse p
Triggers simple or pattern pulse according to set-
tings.
Reset_config Reset
Resets configuration to product default, will
cause EEPROM CRC error on next boot.
reset
Reboots the board, maintains most settings.