December 2019  28  LDC-3706 Series Laser Controller 
 
Operating a Laser in Constant Current (Current) Mode with Analog 
Modulation 
1.  Plug the LDC-3706 into an appropriate AC power source supplying the correct mains voltage 
and frequency for the instrument (refer to the rear panel for the correct ratings) 
2.  Turn on the LDC-3706.  The laser output will be disabled at power up and the unit will 
automatically configure its parameters to the state which existed when the power was last 
shut off. 
3.  If the temperature controller is required please refer to the section below on temperature 
control prior to enabling the laser current output. 
4. Press the LASER MODE soft button until the CURRENT LED is illuminated.  
5.  Select the maximum output current range for the laser by pressing the RANGE soft button 
under LASER RANGE.  For best performance select the lowest range that allows the laser to 
be run at the maximum desired output current. 
6. Press LASER under PARAMETER to adjust the laser current limit, voltage limit, power limit 
(if applicable), calibration PD value, photodiode reverse voltage bias, and LD TEC error link.  
Also make sure to adjust to either LOW or HIGH bandwidth mode, based on the maximum 
required modulation frequency. The maximum frequency bandwidth for each range can be 
found in the specifications in Chapter 1. 
7.  Press the soft button under MOD to enable the input BNC jack and connect the voltage 
signal that will be superimposed on to the output.  See the Specifications in Chapter 1 for 
details on transfer function and maximum bandwidth. Note that all limits also apply to the 
modulated output, thereby protecting the laser in all operating modes. 
 
Be sure not to exceed the specified modulation voltage level or damage to the 
instrument may result. 
8. Under the ADJUST section press the LASER soft button and use the knob to adjust the 
output current to the desired output.  Press the ON soft button under LASER MODE.  It is 
advised when using a new laser or connection to the laser to set the output to zero prior to 
enabling the output.  Once the output is enabled slowly ramp the current to the desired 
output. 
9.  The user can display measured voltage, photodiode current, calculated photodiode power, 
and current setpoint on either the left or right side of the LASER display by pressing the 
LEFT or RIGHT display soft buttons under the LASER display.