December 2019 86 LDC-3706 Series Laser Controller
SOURce[1]:TEMPerature:AM[:STATe] <bool> TEC
SOURce[1]:TEMPerature:AM[:STATe]?
Description Sets/reports TE analog modulation enabled state.
Parameters 0 or 1 – 0 is disabled, 1 in enabled
Reset Value 0
Example SOUR:TEMP:AM 1; Turns on the analog temperature modulation.
SOURce[1]:TEMPerature:LCONstants:DERivative <nrf> TEC
SOURce[1]:TEMPerature:LCONstants:DERivative?
Description Sets/reports derivative term for TEC controller PID loop in temperature mode.
See Chapter 2 for a summary of PID control.
Parameters 0.000000 – 10.000000 – Derivative value setting
Reset Value 0.05
Example SOUR:TEMP:LCON:DER 0.05; Sets the derivative term for temperature mode to 0.05.
SOURce[1]:TEMPerature:LCONstants:INTegral <nrf> TEC
SOURce[1]:TEMPerature:LCONstants:INTegral?
Description Sets/reports integral term for TEC controller PID loop in temperature mode.
See Chapter 2 for a summary of PID control.
Parameters 0.000000 – 10.000000 – Integral value setting
Reset Value 0.10
Example SOUR:TEMP:LCON:INT 0.1; Sets the integral term for temperature mode to 0.1.
SOURce[1]:TEMPerature:LCONstants[:GAIN] <nrf> TEC
SOURce[1]:TEMPerature:LCONstants[:GAIN]?
Description Sets/reports proportional term for TEC controller PID loop in temperature mode.
See Chapter 2 for a summary of PID control.
Parameters 0.000000 – 100.000000 – Proportional value setting
Reset Value 15
Example SOUR:TEMP:LCON 15; Sets the proportional term for temperature mode to 15.
SOURce[1]:TEMPerature:PROTection:LOW <nrf> TEC
SOURce[1]:TEMPerature:PROTection:LOW?
Description Sets/reports low limit for temperature.
Parameters -100.00 – 200.00 – Low temperature limit in ˚C.
Reset Value 0
Example SOUR:TEMP:PROT:LOW 0; Sets the lower limit for temperature to 0ºC.