84 Remote Command Set
ST
Description Stop motion command
Syntax xxST
Remarks This command is used to stop the motion of an axis. The controller uses acceleration
specified using AC command to stop motion. If no axis number is specified, the
controller stops the axis that is currently moving. Use Abort (AB) command to abruptly
stop motion without deceleration.
Argument Value Description
xx 1 to 4 Axis number
Example
1MV+ (Move motor 1 indefinitely in positive direction)
1ST (Stop motion)
ST (Stop motion on any motor that is moving)
See Also AB, AC