Electromagnetic Compliance
Manufacturers declaration of electromagnetic immunity
The ODFS
®
Pace (XL) is intended for use in the electromagnetic environment specified
below. The customer or user of the ODFS
®
Pace (XL) should assure that it is used in such
an environment.
Immunity test IEC 60601
test level
Compliance
level
Electromagnetic environ-
ment - guidance
Electrostatic discharge
(ESD) IEC 61000-4-2
± 8kV contact
± 15 kV air
± 8kV contact
± 15 kV air
Floors should be wood,
concrete or ceramic tile. If oors
are covered with synthetic
material, the relative humidity
should be at least 30 %.
Electrical fast transient/
burst IEC 6100-4-4
No
connection
to mains
(Battery-
operated)
ODFS Pace® (XL) does not
have a mains power supply
Surge IEC 6100-4-5 No
connection
to mains
(Battery-
operated)
Voltage dips, short
interruptions and
variations
IEC 61000-4-11
No
connection
to mains
(Battery-
operated)
Power frequency
(50/60Hz) magnetic
eld IEC 61000-4-8
30 A/m 30 A/m Power frequency magnetic
elds should be at levels
characteristic of a typical
location in a typical commercial,
hospital or domestic
environment.