49
Electromagnetic Compliance
Electromagnetic Emissions
The ODFS
®
Pace was tested with electrodes, footswitch, electrode lead and footswitch lead
of the maximum length (electrode lead – 1.5m, footswitch lead – 1.5m and footswitch –
0.6m). The ODFS
®
Pace XL was tested with 2 OML LINQ
TM
using one 0.6m footswitch in
each and the OML Remote. Details of these accessories can be found in the manual.
If the performance of the system is compromised by other equipment (such as
home routers), the user should turn o the system and move away from the
interfering equipment. Please refer to fault nding.
Like all radio telemetry devices the ODFS
®
Pace XL and wireless accessories
should not be used while flying. Set the ODFS
®
Pace XL to airplane mode and
use a wired footswitch. Please refer to page 27 on how to set airplane mode.
Manufacturer’s declaration of electromagnetic emissions
The ODFS
®
Pace (XL) is intended for use in the electromagnetic environment specified
below. The customer or user of the ODFS
®
Pace (XL) should ensure that it is used in such
an environment.
Emission test Compliance Electromagnetic environment guidance
RF emissions CISPR 11 Group 1
The ODFS
®
Pace (XL) uses RF energy
only for its internal function. Therefore, its
RF emissions are very low and are not
likely to cause any interference in nearby
electronic equipment.
Radiated Spurious
Emissions
ETSI EN 300 328 V2.1.1
Clause 5.4.9
(30MHz to
12.75GHz)
Radiated emissions
FCC/CFR 47:Part
15.209:2017
Class B
(30MHz to
25GHz)
RF emissions CISPR 11 Class B
The ODFS
®
Pace (XL) is suitable for use
in all establishments, including domestic
establishments and those directly
connected to the public low voltage power
supply network that supplies buildings
used for domestic purposes.
Harmonic emissions IEC
61000-3-2
No connection to
mains (Battery-
operated)
Voltage uctuations/
icker emissions IEC
61000-3-3
No connection to
mains (Battery-
operated)
6