System Shutdown
1. Clean any oil immersion lenses used with the lens tissue
2. Close Nikon Nis-Elements
3. If the 640nm laser is turned on, maximise the 640 control software and select Off
then close the window
4. Shutdown the PC
Incubation chamber heater and CO2 supply if used
5. MCL Piezo (on shelf)
6. XY stage controller (on shelf)
7. A1 Controller
8. Prior brightfield LED, Switched off via Prior Optiscan II box
9. Lasers- Turn the key on the front of the laser bed anticlockwise to the vertical
position
Mounting a sample
1. Ensure the correct stage insert is mounted (see facility
staff if unsure)
2. Select a low magnification (10-20x) lens from the
touchpad objective menu
3. Use the joystick to move the stage if necessary.
Change the speed of the stage by twisting the joystick.
4. Before mounting the slide on the stage, turn the microscope focus
knob, if the Z value on the microscopes front LCD display doesn't
change as you turn, the lens is in the Escape position. Press the
Refocus button on the right side of the microscope body.
5. Remember to invert your slide so that the cover glass is the closest surface to the
lens

Other manuals for Nikon A1+

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Nikon A1+ and is the answer not in the manual?

Nikon A1+ Specifications

General IconGeneral
TypeConfocal Microscope
SoftwareNIS-Elements
Laser Lines405 nm, 488 nm, 561 nm, 640 nm
DetectorGaAsP PMT detectors
Scanning MethodGalvanometer-based scanning
Objective TurretMotorized
DimensionsVaries by configuration
WeightVaries by configuration
Scan SpeedUp to 30 fps (512 x 512 pixels)
Objective LensesCompatible with Nikon CFI series objectives
Frame Rate30 frames per second

Related product manuals